EDX Spectroscopy and Mapping
STEM Calibration setup
Adjust it to be CL2 aperture #1
Decide what probe size to use between "4C" and "6C".
The "4C" probe size is brighter and needs a shorter time for enough signals.You may use Orius for calibration; Exposure time: 0.001 for instant image acquisition.
After the A1 and B2 alignment, we shall re-PL align to the detector.
EDX Software
Switch on the AZtec software >> New Project >> Browse >> Map.
Thermal >> Operate >> Cooling >> Cold
The EDX Detector works at -20oC, it is cold down by thermal electric coupling.
Cold Temperature protects the detector and enhances the signal.Scan image >> "Setting AutoSetting": ON >> Drift correction with scan 512; dwell time 10.
1024 is too slow for drift correction.ADF 1, 512 Always Autolock shift --> time 2X, 4X, and 8X buffer.
For high magnification: 8X.
For low magnification: 2X.
View >> MinView: Auto Progression; monitoring the MiniView and adjust the focus if drifted
Drift over time: Sample stage and Focus.You may choose different line Series: K or M to increase the brightness.
When finish the EDX characterizations, remove the detector and set it to Standby mode (Room temp)
A positive x-tilted sample stage can improve the EDX signal.